Tag Archives: transfusion

Blood Warmers (Sort Of) In The Wild

This post (after, let’s face it, a massive break) is the written version of a talk by Andrew Weatherall for the Aeromedical Society of Australasia Conference 2019, held in Perth.

 This is a thing that’s about blood. And the roadside. And the lab bench. Naturally it starts with a story because clinical research should start with a patient. That’s why we do it. It’s a really topic to talk about because the management of transfusion, and the questions it brings up might seem simple at first glance. But a transfusion in the prehospital environment is a particular example of people across a whole system working for a patient.

And the questions clinicians ask that start with something small are the sorts of questions that make a difference in simple ways to every single patient.

A Mountain Story

You have to cast yourself back to the late 90s. Back when the millennium bug was a spooky tale and not an impressively well managed systems vulnerability.

On a particular morning back then there was a car accident in the Blue Mountains, just outside of Sydney. It was serious. The driver had died.

These were the times when asking for a team with a few more options waited until the ambulance officers had got to the scene and started working very hard. So around 28 minutes after the accident, the CareFlight crew on that day got a call to attend to try and help a passenger. Around 52 minutes had passed by the time they reached the patient.

In that time a lot of difficult treatment and rescue work had already happened. The patient, a 15 year old was accessible but still a bit stuck. The paramedics had already started care. In this case this included a large bore cannulae and 6 litres of polygeline fluid resuscitation.

I should have mentioned it was 7 degrees Celsius this particular morning.

Over the course of the extrication all the crews on scene combined. The patient was intubated because the combination of their injuries, diminishing conscious level and the respiratory rate of 40/min suggested something was up.

They received the 4 units of red cells that the CareFlight crew was carrying. The heart rate of 140/min and blood pressure of 70 mmHg made it clear bleeding was an issue. It was an issue that needed more red cells. How does that happen on a cold day in the mountains in 1997? The local hospital, nothing like a trauma hospital sends O negative red cells when they receive a call. The police deliver it.

The patient had a chest drain inserted after their breath sounds changed unilaterally.

But it was clear they needed more fluid resuscitation. And more.

In total, by the time the patient reached their destination hospital a flight later and about 132 minutes after their injury, the combined teams had managed to find a total of 15 units for that patient. Which gave the hospital a chance to get working.

The receiving hospital that day was Nepean hospital, at the foot of the mountains (and these days not a major trauma centre). They received a critically unwell patient and had a long day in the operating theatres. A long day with a long list of injuries who received a further 56 units of red cells, 16 units of platelets and 19 units of FFP.

And survived.

I know they survived because they provided consent for the first case report of massive transfusion in the prehospital setting that Rob Bartolacci and Alan Garner wrote up in the MJA in 1999.

Questions

That story could be seen as an extraordinary success. It didn’t stop the crew asking questions though. This was of course before we knew nearly as much about trauma and bleeding and coagulopathy. I mean we knew it was an issue but Brohi et al hadn’t published that excellent work showing a rate of coagulopathy of 24.4% or the huge import of coagulopathy when it came to mortality.

We didn’t have nearly as much understanding of the ills of acidosis, coagulopathy and hypothermia. We didn’t even have as much evidence about how easy it is to cool patients with cooled fluids as we now do.

We knew hypothermia was bad though. So when that patient arrived to hospital with a temperature of 29.5 degrees Celsius and a heart rate of 80/min, the crew started thinking about ways to do it better.

These days of course we prioritise haemorrhage control. We have a different approach to administering massive transfusions. We’d be reaching for tranexamic acid. Along the way though the first question was ‘how do we try and make our fluids warm?’ Red cells come out of the esky at around 4 degrees Celsius. Patients are not designed to meet that temperature halfway.

Lots of things have been tried over the years. In the past we’d relied on the Australian sun. We’d relied on the toasty armpit of an emergency service worker probably wearing a non-breathable fabric blend that isn’t very flattering to the profile. We’d tried gel heat pads that you’d use to ease a muscle ache.

Fast forward a decade and a bit though and we finally had a portable fluid warming device small enough to help us out. We felt pretty good about the Belmont Buddy Lite too. It was a huge step up, we figured, from what we’d been doing.

Clinicians still ask questions though. This time it was another CareFlight specialist, James Milligan, who started asking questions. It was a pretty simple one really. ‘Maybe we should check how much better it is than all the other options we’ve tried?’

Simple, right?

 

Setting the Right Rules

Now one of the great challenges when you’re trying to do bench testing of a device for prehospital and retrieval medicine is getting the balance right. You want to produce measurements that are rigorous and reliable enough to give you real information. The risk when you do that is that you do things so differently to the environment that counts for us that it no longer represents something that still applies at the roadside.

So the natural choice if you want to bench test a prehospital blood warming device is this …

… a bespoke cardiopulmonary bypass circuit to deliver reliable flows, measure the temperature at multiple sites, measure pressure changes across the circuit, collect the first unit of red cells you administer, spin it down and cool it then recirculate that blood to deliver the equivalent of a second unit.

With a plan to fix the flow rates to 50 mL/min (the suggested rate for the Buddy Lite), randomise the sequence of runs and repeat each type of run 3 times to generate useful data. We standardised as much as we could, including the spine board we used for the ‘on a warmed spine board in the sun’ group and had the board heating for a standardised period of time. We used a single armpit to generate body heating. (Yes, I can confirm that sometimes you have to sacrifice a few frozen off armpit hairs for science.) You get those gel pads ready to go. You get the support in an entirely different setting from your blood bank.

Then it’s time to test how MacGyver works in real life.

 

Phase 1: A Warmer vs MacGyver.

This study turned out to be one of those ones where the results match pretty much what you thought. You can find the full paper here but the key table to pull out is this one.

Even with this simple study there are a few interesting points to note:

  • There is actually more warming than you might expect when you just run the fluid through an intravenous fluid line.
  • The Australian sun actually did pretty well.
  • My armpit is just an embarrassment. Only for this reason.
  • Gel pads just don’t have the contact time to count.
  • The thing specifically designed for warming turns out to be better than warming.

Phew. Done and dusted.

Except we had more questions.

The thing is as clinicians we knew that delivering at 50 mL/min is probably not what we do when a patient is really critically unwell. There’s every chance it’s quicker. It certainly was in 1997. Maybe even pulsed because someone is squeezing it in.

So to really make the benchtop more like the place we work, we wanted to test different flow rates. We also thought maybe we should double check that putting red cells through a rapid temperature change across a small area of space where at least some pressure changes happen inside the device was still OK for red cells. What if we haemolyse a bunch of hose bendy little discs.

Happily just as we got there, more devices hit the market.

 

Phase 2: Warmer vs Warmer

We had some more limits to set though. The devices we’re after for prehospital work need to be light enough to not be too annoying. They have to be free of an external power source. Ideally they’d be idiot proof. I mean folks like me just a little overloaded at the accident scene need to use them.

So we looked for devices that could do the job at a total weight under 1 kg and that operated as standalone units. We ended up with 4 to test:

  • The Thermal Angel.
  • The M0
  • The Hypotherm X LG.
  • The Buddy Lite.

This time with flow rates of 50 mL/min, 100 mL/min, and 200 mL/min (the maximum rating of any device was 150 mL/min but we also knew that with a pump set we can get up as high as those numbers). Still randomised. This time with fresh units of red cells so we could test for haemolysis on the first run through the circuit.

The testing circuit was a little different this time, mostly because it did things better and would reliably deliver the flow rate all the way to 200 mL/min. It was a bit quicker to cool the units too.

The results this time are best shown in a couple of the tables. At 50 mL/min there’s one warmer that’s clearly performing better than the others. The red cells reach the thermistor that reflects delivery to the patient at 36.60C. The best of the others only gets to 30.50C.

When you get to higher flow rates the difference is even more marked.

That same device gets the blood to 32.50C. The others? 23.7, 23.5 and 19.40C.

That is a heck of a difference.

Along the way we couldn’t pick up any evidence of haemolysis. We count that as the best sort of negative result.

Oh, that warmer was the M warmer. We switched.

The Deeper Bits

A study that started with a very simple question ended up being a pretty fun excursion into the lab. Except for that one thing where the clamp went on the wrong bit of the circuit.

For no reason at all I’m just going to mention that wearing PPE not your own clothes is something you should be glad you do at work. It pays off.

The other thing worth noting is that all of these devices go through a process before they become available for sale commercially. That is part of what the TGA does. And they all do kind of what they say on the box. They do warm things up. They are unlikely to cause problems. They’re not the same though. You could even argue that some of them performed so below the level of the most effective device we tested that it’s only marginally better than our next reference, my armpit.

Everything that goes to market has its own story, just like the little pig who goes to market I guess. The TGA is obviously very rigorous in applying its incredibly voluminous guidelines. And I wouldn’t suggest that the manufacturers or those sponsoring the devices t get into the market haven’t likewise followed every part of that process.

What is less clear to me is how you demonstrate ‘suitable for intended use’ which is one of the essential principles that must be shown to be met. Near as we can tell there wasn’t prior testing of these units that so rigorously reflected how we’re likely to use the units in our actual practice prehospitally.

There is a requirement to show that the design meets appropriate standards and that there is clinical evidence. However you are permitted to show that testing is similar enough to the area you reckon it can be used that it gets signed off. On my read of the guidelines you can even show good evidence for the principle of blood warming and that a similar device has been shown to do that well and say ‘our tech specs show we can do the same thing so tick here please’. Extrapolating practice from the hospital or other settings to prehospital and retrieval medicine is a thing we often have to do but that doesn’t mean we should just accept that.

So when you look at a design like, say, the Buddy Lite (a device that has served us really well and that was a huge step up from what we were doing) you can see that it would meet lots of essential principles about not exposing the patient to harm, being safe to handle and operate and warming the infusate. As long as it is delivered as suggested by the manufacturer at 50 mL/min.

And that’s the bit of evaluation that we can’t rely on the TGA to do for us. Is 50 mL/min what we’re after?

We need clinicians asking questions.

Slow Down

So we started with a question. Actually the questions have been happening since at least the 90s.

Simple questions asked by clinicians thinking about the patient in front of them are really useful. They take you in unexpected directions. They lead you to work in teams just like you do at the accident scene. It’s just that this team involves some doctors and a paediatric perfusionist and a haematologist and a lab technician and the blood bank and the Red Cross and a statistician in Hong Kong.

And we’ll keep asking questions. Along the way on this study we figured out that the Hypotherm was just a little challenging to use. In using the M warmer we’ve picked up ways we need to manage the battery recharging.

Clinical teams are vital to doing things better for our patients not just because we actually do the doing, but because the studies that get out there need to be interpreted by people who know the operating space. Or it needs to be clinical teams and patients inspiring studies in the first place.

Someone needs to keep an eye on the pigs when they’re trying to get to market.

 

The References Bit:

OK that’s quite long.

That case report is this one (and obviously the patient has given permission for its use in contexts like this):

Garner AA, Bartolacci RA. Massive prehospital transfusion in multiple blunt trauma. Med J Aust. 1999;170:23-5.

The first of the blood warming papers is this one:

Milligan J, Lee A, Gill M, Weatherall A, Tetlow C, Garner AA. Performance comparison of improved prehospital blood warming techniques and a commercial blood warmer. Injury. 2016;47:1824-7.

The more recent one comparing devices is this one:

Weatherall A, Gill M, Milligan J, Tetlow C, Harris C, Garner A, Lee A. Comparison of portable blood-warming devices under simulated prehospital conditions: a randomised in-vitro blood circuit study. Anaesthesia. 2019;8:1026-32.

If you want to read that Brohi et al paper again it’s here:

Brohi K, Singh J, Heron M, Coats T. Acute traumatic coagulopathy. J Trauma. 2003;54:1127-30.

You might like to reflect on just how quickly you can cool someone down with cold crystalloid:

Kämäräinen A, Virkkunen I, Tenhunen J, Ali-Hankala A, Silfvast T. Induction of therapeutic hypothermia during prehospital CPR using ice-cold intravenous fluid. Resuscitation. 2008;79:205-11.

If you have lots of spare time you might like to read the TGA regulations. (Note they are being updated.) I advise drinking coffee first.

And did you get all this way? Then you definitely need to watch this and relax.

 

Making Things Clot

It’s been a solid break but here is Andrew Weatherall returning with a bit about just one approach to getting access to plasma in the field. 

In prehospital medicine there aren’t really many limits to what we can carry. Ultrasounds are routine. Intubation gear is boring. We even carry patients some of the time.

Plus since 1987 we’ve carried red cells for transfusion. Which is excellent. And we’ve used plenty. Sometimes though you want to change things up and changing things takes time. Like say, changing what you carry for transfusion. Which only took us 30 years or so.

 

How you get it done…

Us carrying red cells has been reality through a range of different attitudes to transfusion. Where we find ourselves now of course is thinking that if you need fluids in the setting of trauma then the fluid should not be see through. For a while though we’ve been trying to figure out how to extend beyond red cells.

The thing is that now that we know that acute traumatic coagulopathy (ATC) is a thing, and we’ve known that for a pretty decent amount of time now, you need to be wrestling with how to try and deal with the coagulopathy bit, not just dish out the vampire-sating stuff.

 

So what to use…

If you go to look at the literature, for a while there’s been plenty out there saying that fibrinogen might be a useful thing to go with. So we’ve been waiting a while for freeze-dried cryoprecipitate but last time we looked it came out as a little bit pricey for our taste but there was something more significant.

Time.

djim-loic-69263-unsplash
Blurry, blurry time.

The more significant thing for our work profile was just the fact that preparing FCCs takes a bit of time and would probably take a team member out of action for too long. That might be a different story if we were doing retrieval or more remote pick-ups, but for metropolitan responses it just didn’t quite seem to fit.

So we kept giving the red cells. Then when TXA gained some evidence, we added that. That one’s quick.

Trauma centres kept publishing on the role of viscoelastic testing but that didn’t fit our work profile either. So we were stuck there.

And whole blood was not on the table in our neck of the woods.

It kept nagging at us though. What’s the other thing we should be giving?

Which is why we went to chat to a clever person at the place we pick up our blood.

 

The Ideas Bank

At this point something we’d discounted as an option suddenly became an option. Plasma.

It just hadn’t seemed like an option for us because of the issues around thawing time and the need to use it within 4 hours. That’s not true anymore though. Extended life plasma (ELP) is now available and is OK to use for 4 days after thawing. Add that to some evidence that group A ELP is suitable for use as a donor unit in the adult patient population (rather than having to always use group AB) and our blood bank contact had come up with a plan to do something different.

So this was the plan:

  • Switch what was in our blood esky from 3 units of O- red cells to 2 units of O- red cells and 2 units of group A ELP.
  • At the end of the shift (this operation is running one shift per day), return the products as usual to the blood bank.
  • The ELP can then go back into the pool used by the hospital blood bank.

And we figured we’d start the plasma early, maybe after the first lot of red cells.

There were to be some limits of course, particularly when it comes to kids. Try as we might, we just couldn’t find guidance for kids under 11 to see if group A ELP alone would be appropriate. Same story for the paeds haematologists. So kids were out.

 

The Introduction

The whole thing got moving within a few short months and the plasma hit the esky in April. It was introduced with lots of discussion and education and an undertaking to keep track of the results.

 It really was seamless. We just sort of needed the evidence to back up the expert advice of the haematologists.

 

Catching Up

And this is where another bit of fortuitous timing worked in our favour. A trial came out.

PAMPer reared up in the NEJM showing a very solid difference in mortality in those given plasma in the prehospital phase. This was patients between the age of 18 and 90 and with evidence of haemodynamic things that matter – systolic blood pressure under 90 mmHg (or a single reading hitting 70 or below) and heart rate above 108/minute. 5 minutes of being arrested was also enough to get a number of people kicked to the side.

It’s also worth mentioning that they just gave their 2 units of plasma in their entirety. Some of the centres were carrying red cells so if they were being treated as recruitment centres administering “standard” resuscitation for a particular day, there was a total of 13 of the 27 centres who might go ahead and give some red cells.

A total of 501 participants were eventually randomised. 30 days after the patients were randomised the number of deaths in the standard care group was 89. This is significant compared to the 53 deaths in the plasma group. After regression analysis, those in the plasma group had a risk of death 39% lower than standard care. There were also mortality differences at 24 hours and in the “in-hospital” cohort.

As a bonus, patients in the plasma group received fewer units of blood components overall. Plus they had no documented cases of transfusion-related lung injury during the trial.

So that was enough to make us tweak things a little more and consider the plasma as our first line resuscitation in most instances, followed by the red cells.

 

What makes it work?

So for now this solution works in our context. That’s not to say it would work for everyone or would be the best choice for different work profiles. We note that London’s excellent HEMS service has just recently introduced bags with both red cells and FFP in them and that’s a different approach again which will be worth watching. And even more recently the crew from Kent Surrey Sussex Air Ambulance Trust have published a retrospective study sharing their success with freeze-dried plasma.

It’s also not to say we won’t seek to tweak things in the future. There are a few key things that made the use of ELP work for us:

  1. Local experts

There’s no doubt that the ‘who you know’ factor helped in setting this up. Being able to approach the local expert, in our case Dr Leo Pasalic (full credentials below), who could think laterally when we approached with a query opened up an avenue we had dismissed. Leo was also able to place it in the context of a broader understanding of the literature. There’s no literature search that’s a replacement for long experience and cultivated expertise.

  1. A very excellent blood bank

A nearby blood bank that supports that expert person with great service is a pretty crucial part of the mix. As clever as Leo is (and he is pretty clever), the whole plan hinges on other clever people who pull off the ‘let’s make this happen’ bit and have that extra understanding of the science and logistics. For us those other key people were Dr Kifah Shahin and Hayley Keenan (you should check their full credentials below too). From the time of deciding to do this to actually carrying the ELP was weeks at most and the logistics have never been an issue. They added the packs and made it work.

  1. Reduce, reuse, recycle

When giving blood products we are, naturally, also trying to nourish our inner greenie. I mean, as it is we’re aiming to reduce the overall usage of blood products throughout the admission. We also want to make sure we’re not thawing the ELP and not being able to use it. Blood products are a precious resource and the cost of production is also high. To date, not a single unit of ELP has gone unused. It’s thawed for the prehospital service, returned at the end of shift and invariably used well within the permitted timeframe by the very big hospital for other patients.

gary-chan-351213-unsplash
This is a terrible place for blood products to end up.
  1. The follow-up plan

The plan for tracking the experience of use is already in place so we can confirm our local experience with this new option. Trial evidence is great but ongoing real world monitoring has to come with the territory.

We’ve come a fair distance from our early forays into prehospital blood administration which involved the side of a road, a hospital somewhere nearby and police doing high speed runs to deliver those units.

It’d be good if the next step took less than 30 years though.

 

Notes:

Before we get into the references, we would really like to acknowledge the support of the NSW Health Pathology through their Blood Bank at Westmead Hospital and the brilliant staff working there.

Dr Leo Pasalic is a staff specialist in laboratory and clinical haematology and transfusion medicine and gets to cover the medical supervising responsibilities at the Blood Bank at Westmead Hospital.

Dr Kifah Shahin is the Transfusion Lead Scientist for the West, Rural and Regional laboratories in NSW Health Pathology.

Hayley Keenan is a Transfusion Senior Scientist at NSW Health Pathology Westmead.

That Blood Bank they work at is part of the Department of Haematology at the Institute of Clinical Pathology and Medical Research, NSW Health Pathology, Westmead Hospital. The hospital has been on that location for 40 years. The ICPMR got there first, having been on the site for 41 years.

Now, the reading list.

Here’s the link to the PAMPer thing again:

Sperry JL, Fuyette FX, Brown JB, et al. Prehospital Plasma during Air Medical Transport in Trauma Patients at Risk for Hemorrhagic Shock. New Engl J Med. 2018;279:315-26.

This isn’t the only story out there of course. The paper from Moore et al is worth a look.

Moore HB, Moore EE, Chapman MP, et al. Plasma-first resuscitation to treat hemorrhagic shock during emergency ground transportation in an urban area: a randomised trial. Lancet 2018;392:283-91. 

And that Kent paper again:

Oakeshott JE, Griggs JE, Wareham GM, Lyon RM. Feasibility of prehospital freeze-dried plasma administration in a UK Helicopter Emergency Medical Service. Eur J Emerg Med. doi:10.1097/MEJ.0000000000000585

Now, some others:

Winearls J. Fibrinogen in Traumatic Haemorrhage: A Narrative Review. Injury. 2017; 48:230-42.

Lance MD, Ninivaggi M, Schols SEM, et al. Perioperative dilution coagulopathy treated with fresh frozen plasma and fibrinogen concentrate: a prospective randomised intervention trial. Vox Sanguinis. 2012;103:25-34. 

Schöchl H, Nienaber U, Maegele M, et al. Transfusion in trauma: thromboelasfometry-guided coagulation factor concentrate-based therapy versus standard fresh frozen plasma-based therapy. Critical Care. 2011;15:R83. 

Shackleford SA, del Junco DJ, Powell-Dunford N, et al. Association of Prehospital Blood Product Transfusion During Medical Evacuation of Combat Casualties in Afghanistan With Acute and 30-Day Survival. JAMA. 2017;318:1581-11.